User interface overview

Icon in the Windows notification area

The icon in the notification area of the operating system reflects the current QA status of the workstation.

  • Compliant: All applicable QA tests have a compliant QA status. There currently are no tasks due for execution.

  • Awaiting Results: At least one task is due for execution.

  • Not compliant: At least one QA test has a Not compliant QA status because it failed to pass the test criteria.

Status panel and workstation details

The QAWeb interface user interface consists of the following elements:

  • The QAWeb agent status panel displays summarized information to the user of the workstation. Open the status panel by left-clicking the QAweb agent icon in the Windows notification area. See Using the Status panel.